Diagnostic Reference Level Dose in Routine Digital Radiology Examination in Tehran

Document Type : Original article

Authors

1 MSc Medical Physics, Department of Medical Physics and Biomedical Engineering Shahid Beheshti University of Medical Sciences Faculty of MedicineTehran, Iran

2 Ph.D Biophysics, Department of Medical Physics and Biomedical Engineering Shahid Beheshti University of Medical Sciences Faculty of MedicineTehran, Iran

3 MSc Solid State Physics, Department of Physics, Faculty of Physics and Chemistry, Alzahra University, Tehran, Iran

Abstract

Introduction: According to technology development and promoting cultural and scientific development in country and digital radiology equipment, we are seeing replacing those with analog devices, therefore considering that no research has been done in this area we require to prepare DLR for digital radiology. The main purpose of this study was to investigate patient dose in common X-ray examination and to establish DRLs in digital radiography in Tehran.
Materials and Methods: The study was performed on 2135 patients in 2014-2015 years. Diagnostic reference dose level (DRL) has been studied on adult in 15 digital radiology centers in routine digital radiography examinations in Iran. The dosimetry protocol in this study was indirect method to measure Entrance Surface Air Kerma (ESAK). Finally, the third quarter determined diagnostic reference level .
Results: Based on the results of this study, national diagnostic reference levels were established: 1.3 mGy for Skull(PA), 1.17 mGy for Skull (Lat), 0.77 mGy for Cervical spine(AP), 0.85 mGy for Cervical spine(Lat), 0.63mGy for chest(PA), 1.11 mGy for chest (Lat), 2  mGy for abdomen, 2.69 mGy for lumbar spine ( AP), 4.22 mGy for lumbar spine( Lat), 1.73mGy for thoracic spine (AP) and 2.35 mGy for thoracic spine (Lat). 
Conclusion: This article has determined wide variations in radiation dose of x-ray examinations among hospitals in Iran.

Keywords


1.     Hall EJ, Brenner DJ. Cancer risks from diagnostic radiology. Br J Radiol. 2014; 81(965):362-7.
2.     National Radiological Protection Board (NRPB). Living with radiation. London ,UK: Stationary
Office Publication Center. Available at: URL: https://www.gov.uk/government/collections/national-radiological-protection-board-nrpb-report-series; 1992.
3.     Farzaneh MJ, Shandiz MS, Vardian M, Deevband MR, Kardan MR. Evaluation of image quality and patient dose in conventional radiography examinations in radiology centers in Sistan and Baluchestan, Iran and comparing with that of international guidelines levels. Indian J Sci Technol. 2011; 4(11):1429-33.
4.     Carmichael JH. European guidelines on quality criteria for diagnostic radiographic images. London: Office for Official Publications of the European Communities; 1996.
5.     The Community Research and Development Information Service. Radiological protection and safety in medicine. London: ICRP Publication; 1996.
6.     Edmonds KD. Diagnostic reference levels as a quality assurance tool. Radiographer. 2009; 56(3):32-7.
7.     Protection R. 109: Guidance on diagnostic reference levels (DRLs) for medical exposures. Luxembourg: European Communities; 1999.
8.     Institute of Physical Sciences in Medicine (IPSM). CoR. National Protocol for Patient Dose Measurements in Diagnostic Radiology. London: IPSM; 1992.
9.     Rehani M, Kaul R, Kumar P, Berry M. Does bridging the gap between knowledge and practice help? Example of patient dose reduction in radiology. J Med Phys. 1995; 20(2):18-22.